error MAAP solved

Inviato da Angela M il

flow measured at 24.5 lpm ambient

pressure measured by internal sensor = 11404 hPa

recalibrated to 791 hPa

flow measurement = 16.5 lpm

Flow calibrated at 16.67 lpm

error: 0000 0000 0000 8000 (filter change condition already met immediately after filter change)

flow deviation >5% disappeared

filter changed manually

staus 0000 0000 0000 0000

 

strumento
operatore
data evento
10-Mar-2022
è uno spostamento?
no