Nefelometro TSI 3563 cimone
logbook
Titolo | data | operatore | Descrizione |
---|---|---|---|
fault chopper and shutter | Marinoni Angela | not solved |
|
Calibration and reconnection of internal temperature and pressure sensor | calibrazioni: BLU: old--> k2 7.83 e-4 k4 5.19 e-1 NEW --> k2 7.57377 e-4 k4 5.0904 e-1 GREEN: old--> k2 7.799 e-4 k4 4.98 e-1 NEW --> k2 7.67831 e-4 k4 5.1729 e-1 RED: old--> k2 4.868 e-4 k4 4.86 e-1 NEW--> k2 4.97538 e-4 k4 5.43974 e-1 |
||
cleaning and calinbaration | Marinoni Angela | Pulizia tentativo di calibrazione non riuscito, costanti k4 simili, costanti k2 diverse di 3 ordini di grandezza. |
|
new flow regulation + dryer | Marinoni Angela | set at b15 to have a flow = 15 lpm new dryer on the line, T/RH measured
|
|
pulizia e calibrazione | Pulizia calibrazioni: BLU: old--> k2 8.399 e-4 k4 5.82 e-1 NEW --> k2 7.82987 e-4 k4 5.1861 e-1 GREEN: old--> k2 8.934 e-4 k4 5.56 e-1 NEW --> k2 7.79939 e-4 k4 4.98514 e-1 RED: old--> k2 4.261 e-4 k4 3.41 e-1 NEW--> k2 4.86762 e-4 k4 4.86413 e-1 |
consumabili per strumento
Nome consumabile | marca | modello | note |
---|---|---|---|