installed in TROPOS MPSS
After TROPOS calibration, we decided to move the 3772 into the SMPS, in order to avoid manual correction of data, now necessary, because all the internal corrections were disabled during workshop.
After TROPOS calibration, we decided to move the 3772 into the SMPS, in order to avoid manual correction of data, now necessary, because all the internal corrections were disabled during workshop.
Reinstalled on April, 30.
New T sensors, ok. Spare part, also working in Bologna.
Problem with acquisition, print format?
Installed after WCCAP on April 30 (Cecilia Magnani).
On May 3, full calibration done
SPAN: Ch1 10.711, Ch2 23.046, Ch3 43.421
stability: Ch1 96.544, Ch2 96.517, Ch3 98.002
Zero Ch1 0.1174, Ch2 0.2371, Ch3 -0.178
From 1 second to 1 minute
Nozzle blowed, counting very little. Problem not solved.
FLow measured at 0.33 lpm
The external T probe is broken.
Removed with pump, sensor, cables.
TROPOS intercomparison 22-26
cleaning chamber
flow check
calibration ok, with new following setting saved into the neph:
channel 1 : span 10.001, zero -1.55, stability 97.088
channel 2: span 24.623, zero 0.657, stability 96.756
channel 2: span 45.582, zero 0.4547, stability 96.852
Zero frequency and duration to be changed
1 h measurements with Hepa filter at 0.1 SS from 11 UTC
Zero measurement since 10:27 UTC to 14:00 UTC
misura flusso senza intervento: 16.9 lpm
Pulizia cella di misura. Presenza di tre mosche ai lati del filtro.
Riaccensione, errore pressione. FLusso misurato 19.8 lpm.
Taratura P sensor 11KPa --> 0.787 kPa
Flusso misurato: 16.5 lpm
Calibrazione flusso a 16.67 lpm
zero and start measurements MAAP, ok
already on ozone.isac.bo.cnr.it
fill butanol and silica gel change
fill butanol and silica gel change
Start 11:45 (9:45 UTC) Stop 11:50 (9:50 UTC)
Aerosol generation with latex 1 um not dried.
Nice peak at 1um, plus big peak at 5 mm (hydrated aerosol??)
Filter tape replacement after 5 days without measurements.
ok (10 UTC+1)
Spento in via cautelativa per presenza acqua sulla superficie superiore, infiltrata attraverso il lato esterno della testa di prelievo.
Ricollocato in osservatorio dopo campagna su Aerolab, riparazione presso TSI (sostituzione laser).
presso i laboratori TSI è stato aggiornato il firnware.
Da verificare diametro con PSL
New version of scan software installed (Scan_80).
APS installed and communicate with SMPS laptop (settings: correlated --> summed, Scan single, Baud rate: 9600)
HV calibration solved the problem of the limitation of the first three channels)
Pulizia
tentativo di calibrazione non riuscito, costanti k4 simili, costanti k2 diverse di 3 ordini di grandezza.
Fault su sensore T/RH
set at b15 to have a flow = 15 lpm
new dryer on the line, T/RH measured
PSL 20onm in MIlliQ water, nebulized with pari compact.
Peak seen at 199 nm. Picture attached
11:16 - 12:33 UTC
zero after 1.5 scan, OK
latex 200 nm --> 197 nm (nessuna modifica sheath flow)
test HV fino a 800 V, ok
verifica serosol flow, ok
verifica sheath flow, ok
Riavvio dopo lavori edili
cambio orario UTC+1 --> UTC
Riavvio dopo lavori edili
cambio orario UTC+1 --> UTC
Verifica flusso (3.66 lpm contro i 5 letti dal MFC interno) STD cond.
Flow verification --> flow cal recommended
Flow cal performed and new settings saved. Restarted, no possible to reach the ideal flow at 5 lpm.
New flow set at 4 lpm NIST (0°C, 1013 hPa).
DATA download.
Soluzione del problema: connettore alimentazione pompa non ben inserito, tensore del filtro lento.
Continua a non acquisire
Intervento in loco di Giorgio Tranchida
Installazione PC NUC e test trasmissione dati con TROPOS.
Misura ferma da metà settembre per errore "pump current too high"
Sostituito filtro nel circuito di ricircolo interno, prelevato da GRIMM 1108 "valigetta portatile".
ok
Prelievo porzione di filtro per analisi MWAA e ripristino impostazioni per il cambio filtro da
- 24 h
a
- Trasmittanza <30% e 100 h max
Riavvio dopo lavori edili
cambio orario UTC+1 --> UTC
ciclone rimosso prima dell'interruzione
All'accensione problema sensore di pressione, risolto con calibrazione sensore e flusso (16.67 lpm)
Riavvio dopo lavori edili
cambio orario UTC+1 --> UTC
ciclone rimosso prima dell'interruzione
Riavvio dopo spegnimento per lavori edili,
cambio orario da UTC+1 a UTC
Cancellati dati fino al 1/12/2021 per liberare spazio sul disco.
I dati cancellati sono presenti su ozone.
flow ratio linea B errato (20)
Chiara sottostima delle concentrazioni (<< colonna A)
Mancava idratazione. Software riavviato. Ok
disconnesso dalla linea per installazione cilclone PM1
tra le 15:10 e le 15:20 (UTC+2)
regolazione flusso a 11.67 lpm, installazione ciclone PM1 in comune con AE-33 per campagna ACTRIS/EMEP e misure MWAA.
tra le 15:05 e le 15:25 (14:10 UTC+1)
pulizia camera e sensori, sostituzione filtro
di nuovo errore flusso sensore di pressione
BLU: old--> k2 8.36151 e-4 k4 5.1171 e-1 NEW --> k2 8.39011 e-4 k4 5.81793 e-1
GREEN: old--> k2 8.28784 e--4 k4 5.14255 e-1 NEW --> k2 8.93371 e--4 k4 5.56225 e-1
RED: old-->k2 5.17783e-4 k4 4.95570 e-1 NEW--> k2 4.26085 e-4 k4 3.41276 e-1
check reason for outliers in upper size distribution part
flow measured at 24.5 lpm ambient
pressure measured by internal sensor = 11404 hPa
recalibrated to 791 hPa
flow measurement = 16.5 lpm
Flow calibrated at 16.67 lpm
error: 0000 0000 0000 8000 (filter change condition already met immediately after filter change)
flow deviation >5% disappeared
filter changed manually
staus 0000 0000 0000 0000
0000 0000 0100 8000
codice di errore 0000 0000 0100 8000 0A0100
Problema sul sensore di pressione, completamente fuori range.
calibrazione sensori, calibrazione flusso, pulizia camera e sensori.
Cambio filtro.
status ok, rimesso in misura
Restart con nuovo CPC 3750. Nuova curva di efficiency:
100 1
40 0.987
30 0.993
dalle 10:10 (UTC+1) test filtro, HV off, a zero dopo 2 scan (filtro), subito (HVoff)
rimozione CPC per intercomparison con 3750
Valori crollati a zero in circa 1h.
Livello butanolo ok, flusso ok.
Da verificare
Riavvio osservazioni dopo il fermo per i lavori edili all'osservatorio
Stato strumento:
flusso solo su canale B, flow ratio errato
Conteggio a zero su entrambe le colonne a tutte le SS.
Errore risolto con il riavvio del software di gestione
Installazione, acquisizione su raspberry.
I dati non sono sul server!
il riscaldatore era disconnesso dalla presa di corrente
Tolto ghiaccio esterno alle ore 11:15 UTC+1
Strumento collegato a inlet specifico in rame
Riavviato strumento spedito il 22 ottobre da Bologna
Zero test perfomed, ok
T sensor inlet non working, T red at 0 K
Spare part not available
Pulizia floacked paper
calibrazioni:
BLU: old--> k2 2.413 e-3 k4 5.100 e-1 NEW --> k2 2.54017 e-3 k4 5.15131 e-1
GREEN: old--> k2 2.504 e-3 k4 5.006 e-1 NEW --> k2 2.51907 e-3 k4 5.02744 e-1
!!!ATTENZIONE!!!
Errore nell'impostazione oario del PC
ORARIO impostato su UTC anziché UTC+1 dalle 14 del 4/11/2021 alle 10:55 (UTC+1) del 19/11/2021
Quando si effettua un cambio di orario dal PC è necessario riavviare il programma affinchè le modifiche vengano salvate.
Zero test OK
Aerosol flow: 1.2 lpm (verifica in corso)
Sheath flow: 4.3 lpm
LATEX 200 nm @ 200nm, OK
Pulizia floacked paper
calibrazioni:
BLU: old--> k2 8.374 e-4 k4 5.28 e-1 NEW --> k2 8.36151 e-4 k4 5.1171 e-1
GREEN: old--> k2 9.706 e-4 k4 5.0744 e-1 NEW --> k2 8.28784 e--4 k4 5.14255 e-1
Riavvio osservazioni interrotte per lavori edili
causa inizio lavori edili
causa lavori edili all'osservatorio
Neutral Density Test con kit N AE33-ND-0240 (da Monte Cimone)
ripetuto più volte, per errore sul filtro 1
flow verification (non superato)
--> flow calibration OK con flussimetro TSI e aethalometer PAD (Monte Cimone)
Strumento riportato a Monte CImone dopo riparazione TROPOS.
verifica calibrazione HV, ok
0V --> 0V
4mV --> 2.57V (fuori retta, già visto in passato)
8mv --> 7.5 V
80mV --> 97.3V
800mV --> 1000V
CHeck PSL: picco a 232 nm (P 785 hPa, T=19°C), aggiustato con calibrazione sheath flow
intercetta impostata a 1.16
FLusso sheath misurato dopo settaggio = 3.7 Lpm
cambio sonde T
codice errore 0000 0000 0000 0000
ok
AM + LDL
imballato e spedito il sistema a TROPOS con tutti gli accessori ESCLUSO il CPC
AM + FPC + LDL
portati al CMN un DMA del vecchio sistema MMPS finlandese(?) per capire se il problema potrebbe essere causato da "bruciature" causate da scariche interne al DMA --> anche con il nuovo DMA non si risolve il mistake
portato il FUG "Rinaldi" per provare a generare la HV con un sistema diverso ma le connessioni del FUG non sono concordant con quelle di SMPS--> non si e' eseguita questa procedura
test #3 con componenti SMPS aerolab (AM + LDL)
Montato su SMPS CMN il DMA del sistema aerolab --> per le particelle coarse permane il problema degli alti conteggi
in skype con Kay fatti test su varie componenti per valutare eventuali perdite nel circuito flussi (e.g. nafion) senza ottenere alcun risultato
imballato il sistema e portato a bologna per spedirlo a TROPOS
pulizia floacked paper
calibrazioni:
BLU: old--> k2 1.16 e-3 k4 4.78 e-1 NEW --> k2 8.374 e-4 k4 5.28 e-1
GREEN: old-->k2 1.195 e-3 k4 5.61 e-1 NEW --> k2 9.706 e-4 k4 5.0744 e-1
RED: old--> k2 9.60e-4 k4 6.00 e-1 NEW--> k2 4.8595 e-4 k4 4.5788 e-1
errore 0000 0000 0000 A000
Filter tape fissure: no filter tape transport has been recognized
+
Filter change condition already met again immediately after a filter change has been carried out
Dopo manutenzione pompa, l'errore non cambia. Necessario intervento.