logbook

Titolo data operatore Descrizione
Maintenance Renzi Laura
  • 13:22 UTC Disconnected from the line and turned off 
  • Cleaning of insect trap and optical chamber
  • optical calibration (acceptable)
  • flow calibration, flow measured after calibration 4.980 lpm in STP (stp factor used 0.74)
  • 13:50 UTC reconnected to the line
General Mantainance Renzi Laura
  • Disconnected from the line 08:41 UTC.
  • Cleaning of the optical chamber.
  • Flow calibration.
  • Leakage test at 5 lpm. Result 0% flow difference. Acceptable.
  • Stability test. Acceptable.
  • Clean air test. Acceptable.
  • Flow measured 5.004 lpm (std 101325 Pa, 0°C)
  • Zero performed 12:20-13:40 UTC (Results in table).
  • Reconnected to the line 15:00 UTC. 
acquisition frequency change Marinoni Angela

From 1 second to 1 minute

re-installation after TROPOS workshop

Operatore Cecilia Magnani

Chamber cleaning and disconnection

At 9:30 disconnected to be brought to leipzig for intercomparison. Cleaned after disconnection. 

 

data exported from memory card Marinoni Angela

already on ozone.isac.bo.cnr.it

filter tape replacement Marinoni Angela

Filter tape replacement after 5 days without measurements.
ok (10 UTC+1)

spegnimento cautelativo Marinoni Angela

Spento in via cautelativa per presenza acqua sulla superficie superiore, infiltrata attraverso il lato esterno della testa di prelievo.

flow cal Marinoni Angela

Verifica flusso (3.66 lpm contro i 5 letti dal MFC interno) STD cond.
Flow verification --> flow cal recommended

Flow cal performed and new settings saved. Restarted, no possible to reach the ideal flow at 5 lpm.

New flow set at 4 lpm NIST (0°C, 1013 hPa).

DATA download.

nuovo software ACTRIS Marinoni Angela

Installazione PC NUC e test trasmissione dati con TROPOS.

riavvio e cambio orario Marinoni Angela

Riavvio dopo lavori edili
cambio orario UTC+1 --> UTC

riavvio e cambio ora Marinoni Angela

Riavvio dopo lavori edili
cambio orario UTC+1 --> UTC

ciclone rimosso prima dell'interruzione

spegnimento per lavori edili Marinoni Angela
installazione ciclone PM1 Marinoni Angela

disconnesso dalla linea per installazione cilclone PM1

tra le 15:10 e le 15:20 (UTC+2)

cambio sistema di acquisizione

Acquisizione modificata da psoftware Labview su PC a raspberry, files nella cartella /home/cimone/data/aethal33_dry
NOME FILE:
aethAE33_CMN_YYYYMMDD_p01