MAAP_Cimone

marca
Thermo Scientific
modello
MAAP 5012
inventario ISAC
1076538
osservatorio
assegnazione
Angela Marinoni
data presa carico
tipo misura
misura
stato
funzionante
stato di attività
In misura

ultima posizione

posizione data evento Desrizione
in osservatorio

Reinstalled on April, 30.
New T sensors, ok. Spare part, also working in Bologna.
Problem with acquisition, print format?


 

logbook

Titolo data operatore Descrizione
Maintenance Renzi Laura
  • 15:30 UTC disconnected from the line
  • flow measured ambient condition 17 lpm 
  • cleaned from insects inside the chamber 
  • flow unstable varying form the correct one 17 lpm to 2 lpm periodically
  • 07:55 8-11-2024 UTC reconnected to the line 
  • acquisition issue (some of the settings lost by switching off were not reset correctly, check)
Zero flow Marco Zanatta

flow went to zero at  07:30 UTC...maybe pump dead, to be checked

Zero Marco Zanatta

Zero MAAP  in parallel with MAAP evk2

  • Start time UTC-2 : 12:26
  • EndTime UTC-2: 14:14
Ripartita 09/05/2024 acq. MAAP con correzione print format Calzolari Francescopiero

Ripartita 09/05/2024 acq. MAAP con correzione print format e tempi di send che erano stati cambiati alla calibrazione e manutenzione a Lipsia.

Print format 92

tempi di send 0min; 0secondi     ovvero in attesa di richiesta, nessuna spedizione automatica

richieste fatte dal programma ai 00 secondi di ogni minuto P; #; J2

Reinstalled after WCCAP Marinoni Angela

Reinstalled on April, 30.
New T sensors, ok. Spare part, also working in Bologna.
Problem with acquisition, print format?


 

consumabili per strumento

Nome consumabile marca modello note
filtro a nastro tsi gf 10