Aethalometer AE-33 CMN

marca
Magee Scientific
modello
AE 33-7
serial N
AE33-S08-01045
inventario ISAC
DTA 501.000 – 1075543 -000
osservatorio
prima collocazione
Monte Cimone
assegnazione
Angela Marinoni
data presa carico
tipo misura
misura
stato
funzionante
stato di attività
In misura

ultima posizione

posizione data evento Desrizione
in osservatorio

Operatore Cecilia Magnani

logbook

Titolo data operatore Descrizione
Maintenance Renzi Laura
  • 13:22 UTC Disconnected from the line and turned off 
  • Cleaning of insect trap and optical chamber
  • optical calibration (acceptable)
  • flow calibration, flow measured after calibration 4.980 lpm in STP (stp factor used 0.74)
  • 13:50 UTC reconnected to the line
General Mantainance Renzi Laura
  • Disconnected from the line 08:41 UTC.
  • Cleaning of the optical chamber.
  • Flow calibration.
  • Leakage test at 5 lpm. Result 0% flow difference. Acceptable.
  • Stability test. Acceptable.
  • Clean air test. Acceptable.
  • Flow measured 5.004 lpm (std 101325 Pa, 0°C)
  • Zero performed 12:20-13:40 UTC (Results in table).
  • Reconnected to the line 15:00 UTC. 
acquisition frequency change Marinoni Angela

From 1 second to 1 minute

re-installation after TROPOS workshop

Operatore Cecilia Magnani

Chamber cleaning and disconnection

At 9:30 disconnected to be brought to leipzig for intercomparison. Cleaned after disconnection. 

 