Titolo | data | operatore | Descrizione |
---|---|---|---|
Calibration HV and PSL | Mazzini Martina | è stata svolta la calibrazione HV e PSL 200nm. |
|
reinstalled after WCCAP workshop | Mazzini Martina | MPSS installed, the CPC used is now the TSI_3772, previously used as totCPC. Sheathflow pump installed as compressed air, not in vacuum (as it was before), following WCCAP suggestion. Flow 4.8L/min (ambient)
PSL cal, sheath flow slope re-adjusted from 1.3 (peak 196nm) to 1.2 (peak 200nm) Back to UTC time (UTC+2 used while at WCCAP) Measurment mode ON Next step: grounding DMA |
|
removed for WCCAP calibration/intercomparison | Marinoni Angela | ||
Firnware update, APS connection, HV calibration | Marinoni Angela | New version of scan software installed (Scan_80). APS installed and communicate with SMPS laptop (settings: correlated --> summed, Scan single, Baud rate: 9600)
|
|
PSL check | Marinoni Angela | PSL 20onm in MIlliQ water, nebulized with pari compact. |
|
zero measurement | Marinoni Angela | 11:16 - 12:33 UTC zero after 1.5 scan, OK
|
|
sostituzione DMA | Mazzini Martina | Tolto DMA originale (riparato a Leipzig e attivo da settembre) con DMA in prestito da TROPOS. Verifica sparks nei prossimi giorni |
|
verifica HV e latex | Marinoni Angela | latex 200 nm --> 197 nm (nessuna modifica sheath flow) |
|
cleaning DMA | Marinoni Angela | check reason for outliers in upper size distribution part |
|
cambio CPC | Marinoni Angela | Restart con nuovo CPC 3750. Nuova curva di efficiency: 100 1 40 0.987 30 0.993 20 0.988 14 0.937 11 0.846 10 0.790 9 0.725 8 0.638 7 0.533 6 0.384 5 0.264 |
|
zero test | Marinoni Angela | dalle 10:10 (UTC+1) test filtro, HV off, a zero dopo 2 scan (filtro), subito (HVoff) rimozione CPC per intercomparison con 3750 |
|
Interruzione acquisizione | Mazzini Martina | Blackout Monte Cimone, spegnimento del PC. L'acquisizione si è interrotta da venerdì 17/12 ore 20:30 fino a martedì 21/12 ore 14:30 ca |
|
Rimozione Nafion | Mazzini Martina | E' stato eliminato il nafion dal circuito sheath. Flussi ok. Nafion in riparazione. |
|
Service mode | Mazzini Martina | E' stata avviata la modalità service mode alle ore 11:20UTC+1. La misura riparte alle 14:25UTC+1. |
|
Check vari | Mazzini Martina | Aerosol flow (unchanged configuration) = 1.2 lpm Flow at the DMA IN= 1.2 lpm -Air compressed disconnected form both nafion (pump only to CPC) -Air compressed connected only to CPC and aerosol nafion -Air compressed connected only to CPC and sheath nafion |
|
errore orario PC | Marinoni Angela | !!!ATTENZIONE!!! Errore nell'impostazione oario del PC Quando si effettua un cambio di orario dal PC è necessario riavviare il programma affinchè le modifiche vengano salvate. |
|
check vari | Marinoni Angela | Zero test OK LATEX 200 nm @ 200nm, OK
|
|
riavvio misure SMPS | Marinoni Angela | Strumento riportato a Monte CImone dopo riparazione TROPOS. verifica calibrazione HV, ok 0V --> 0V 4mV --> 2.57V (fuori retta, già visto in passato) 80mV --> 97.3V 800mV --> 1000V intercetta impostata a 1.16 FLusso sheath misurato dopo settaggio = 3.7 Lpm
|
|
manutenzione a TROPOS | Marinoni Angela | AM + LDL imballato e spedito il sistema a TROPOS con tutti gli accessori ESCLUSO il CPC |
|
test # 3 SMPS | Marinoni Angela | test #3 con componenti SMPS aerolab (AM + LDL) Montato su SMPS CMN il DMA del sistema aerolab --> per le particelle coarse permane il problema degli alti conteggi in skype con Kay fatti test su varie componenti per valutare eventuali perdite nel circuito flussi (e.g. nafion) senza ottenere alcun risultato imballato il sistema e portato a bologna per spedirlo a TROPOS
|
|
test # 2 smps | Marinoni Angela | AM + FPC + LDL portati al CMN un DMA del vecchio sistema MMPS finlandese(?) per capire se il problema potrebbe essere causato da "bruciature" causate da scariche interne al DMA --> anche con il nuovo DMA non si risolve il mistake portato il FUG "Rinaldi" per provare a generare la HV con un sistema diverso ma le connessioni del FUG non sono concordant con quelle di SMPS--> non si e' eseguita questa procedura alla fine del test con l'accordo di kay non trovando una soluzione si decide di portare a Bologna l'elettronica del CMN per provarla sul sistema aerolab 30/01/2021 montata l'elettronica del CMN su SMPS aerolab e tutto funziona regolarmente (dopo aver modificato su NI max il device flussimetro che fra i due sistei e' differente)
|
|
TEST #1 smps | CPC replaced --> back to the original |