logbook

Titolo data operatore Descrizione
Calibration HV and PSL Mazzini Martina

è stata svolta la calibrazione HV e PSL 200nm. 

reinstalled after WCCAP workshop Mazzini Martina

MPSS installed, the CPC used is now the TSI_3772, previously used as totCPC.

Sheathflow pump installed as compressed air, not in vacuum (as it was before), following WCCAP suggestion. Flow 4.8L/min (ambient)
HV voltage calibration:

  • 0.4mV:
    • from -0.0039 (1.8v)
    • to -0.0016 (5.2V)
  • 800mV:
    • 0.789 oscillating a bit between 996V to 1002V
  • 200mV:
    • 249V

PSL cal, sheath flow slope re-adjusted from 1.3 (peak 196nm) to 1.2 (peak 200nm)

Back to UTC time (UTC+2 used while at WCCAP)

Measurment mode ON

Next step: grounding DMA

removed for WCCAP calibration/intercomparison Marinoni Angela
Firnware update, APS connection, HV calibration Marinoni Angela

New version of scan software installed (Scan_80).

APS installed and communicate with SMPS laptop (settings: correlated --> summed, Scan single, Baud rate: 9600)
HV calibration solved the problem of the limitation of the first three channels)

 

PSL check Marinoni Angela

PSL 20onm in MIlliQ water, nebulized with pari compact.
Peak seen at 199 nm. Picture attached

zero measurement Marinoni Angela

11:16 - 12:33 UTC

zero after 1.5 scan, OK

 

sostituzione DMA Mazzini Martina

Tolto DMA originale (riparato a Leipzig e attivo da settembre) con DMA in prestito da TROPOS.

Verifica sparks nei prossimi giorni

verifica HV e latex Marinoni Angela

latex 200 nm --> 197 nm (nessuna modifica sheath flow)
test HV fino a 800 V, ok
verifica serosol flow, ok
verifica sheath flow, ok

 

cleaning DMA Marinoni Angela

check reason for outliers in upper size distribution part

cambio CPC Marinoni Angela

Restart con nuovo CPC 3750. Nuova curva di efficiency:

100     1

40    0.987

30    0.993

20    0.988

14    0.937

11    0.846

10    0.790

9     0.725

8     0.638

7     0.533

6     0.384

5     0.264

zero test Marinoni Angela

dalle 10:10 (UTC+1) test filtro, HV off, a zero dopo 2 scan (filtro), subito (HVoff)

rimozione CPC per intercomparison con 3750

Interruzione acquisizione Mazzini Martina

Blackout Monte Cimone, spegnimento del PC. L'acquisizione si è interrotta da venerdì 17/12 ore 20:30 fino a martedì 21/12 ore 14:30 ca

Rimozione Nafion Mazzini Martina

E' stato eliminato il nafion dal circuito sheath.

Flussi ok.

Nafion in riparazione.

Service mode Mazzini Martina

E' stata avviata la modalità service mode alle ore 11:20UTC+1.

La misura riparte alle 14:25UTC+1.

Check vari Mazzini Martina

Aerosol flow (unchanged configuration) = 1.2 lpm

Flow at the DMA IN= 1.2 lpm

-Air compressed disconnected form both nafion (pump only to CPC)
       Aerosol 0.97 lpm

       CPC flow 0.98 lpm
       Flow at the DMA IN= 0.976 lpm

-Air compressed connected only to CPC and aerosol nafion
       Aerosol flow = 1.2 lpm

-Air compressed connected only to CPC and sheath nafion
      Aerosol flow =0.97 lpm

errore orario PC Marinoni Angela

!!!ATTENZIONE!!!

Errore nell'impostazione oario del PC
ORARIO impostato su UTC anziché UTC+1 dalle 14 del 4/11/2021 alle 10:55  (UTC+1) del 19/11/2021

Quando si effettua un cambio di orario dal PC è necessario riavviare il programma affinchè le modifiche vengano salvate.

check vari Marinoni Angela

Zero test OK
Aerosol flow: 1.2 lpm (verifica in corso)
Sheath flow: 4.3 lpm

LATEX 200 nm @ 200nm, OK

 

riavvio misure SMPS Marinoni Angela

Strumento riportato a Monte CImone dopo riparazione TROPOS.

verifica calibrazione HV, ok 

0V --> 0V

4mV --> 2.57V (fuori retta, già visto in passato)
8mv --> 7.5 V

80mV --> 97.3V

800mV --> 1000V
CHeck PSL: picco a 232 nm (P 785 hPa, T=19°C), aggiustato con calibrazione sheath flow

intercetta impostata a 1.16

FLusso sheath misurato dopo settaggio = 3.7 Lpm

 

manutenzione a TROPOS Marinoni Angela

AM + LDL 

imballato e spedito il sistema a TROPOS con tutti gli accessori ESCLUSO il CPC

test # 3 SMPS Marinoni Angela

test #3 con componenti SMPS aerolab  (AM + LDL)

Montato su SMPS CMN il DMA del sistema aerolab --> per le particelle coarse permane il problema degli alti conteggi

in skype con Kay fatti test su varie componenti per valutare eventuali perdite nel circuito flussi (e.g. nafion) senza ottenere alcun risultato

imballato il sistema e portato a bologna per spedirlo a TROPOS

 

test # 2 smps Marinoni Angela

AM + FPC + LDL

portati al CMN un DMA del vecchio sistema MMPS finlandese(?) per capire se il problema potrebbe essere causato da "bruciature"  causate da scariche interne al DMA --> anche con il nuovo DMA non si risolve il mistake

portato il FUG "Rinaldi" per provare a generare la HV con un sistema diverso ma le connessioni del FUG non sono concordant con quelle di SMPS--> non si e' eseguita questa procedura

alla fine del test con l'accordo di kay non trovando una soluzione si decide di portare a Bologna l'elettronica del CMN per provarla sul sistema aerolab

30/01/2021

montata l'elettronica del CMN su SMPS aerolab e tutto funziona regolarmente (dopo aver modificato su NI max il device flussimetro che fra i due sistei e' differente)

 

 

TEST #1 smps

CPC replaced --> back to the original
DMA replaced --> back to the original
CHeck filtered sheth air --> ok
Check aerosol and CPC flow--> ok
HV power supply removed